Not medical advice. A CPAP is a prescribed medical device. Before you rely on any battery or power station, ask your doctor, your equipment supplier, or the CPAP’s manufacturer how your machine should be powered in an outage. The FDA’s guidance for home medical devices says the same thing: read your user instructions or call your distributor or manufacturer to find out whether the device can run on batteries or a generator (FDA, checked Oct 12, 2026).
Short answer: A CPAP is one of the easiest things to back up. Using the figures ResMed publishes in its user guides, a ResMed AirSense 11 plugged into a power station’s AC outlet with its own adapter uses about 450 Wh over an 8-hour night. After inverter losses, that’s roughly 530 Wh from the battery, so a ~1 kWh power station covers about two nights. Turn the heated humidifier and heated tube off and run the machine from a DC source with the manufacturer’s DC cable, and the same battery can last several times longer. ResMed’s own battery tables for the AirSense 10 show about 11 W at a pressure of 10 cmH₂O with no humidifier.
Here’s where those numbers come from, how the humidifier changes them, and estimated nights for the eight power stations we track. To size a CPAP together with a fridge, router, and lights, use the runtime calculator.
Step 1: Find your machine’s power numbers
The best source is your CPAP’s own user guide, usually in the “Technical specifications” section near the back. Here is what the manufacturers publish for the most common home machines:
| Machine | Power supply (as published) | Typical / peak power (as published) | Source |
|---|---|---|---|
| ResMed AirSense 11 | 65 W adapter, 24 V / 2.71 A output | 56.1 W typical / 73.2 W peak | AirSense 11 user guide |
| ResMed AirSense 10 | 90 W adapter, 24 V / 3.75 A output | 53 W typical / 104 W peak | AirSense 10 user guide |
| Philips DreamStation 2 | AC input 100–240 V, 1.0–2.0 A; the manual says an 80 W power supply is required to support humidification | No typical-watt figure in the manual we checked | DreamStation 2 Advanced user manual |
All three documents checked October 12, 2026. ResMed’s “typical” figure doesn’t say which humidity settings it assumes, so treat it as a realistic full-feature planning number rather than a minimum. If your manual lists only a power-supply rating (like the DreamStation 2’s 80 W), use that as a conservative worst case.
Surge isn’t a concern. Even the highest peak above (104 W) is a tiny fraction of the 1,500 W or more that every power station in our dataset can supply from its AC outlets. For a CPAP, capacity (Wh) is the only number that matters.
Step 2: The humidifier is the big variable
The blower in a CPAP uses relatively little power. Heating water and heating the tube use most of it. ResMed’s battery guide (document 198103, checked Oct 12, 2026) lists current draw at 12 V DC for the AirSense 10 when it’s powered through ResMed’s DC-to-DC converter. At a pressure of 10 cmH₂O:
| AirSense 10 setup (ResMed battery guide) | Current @ 12 V | ≈ Watts | ≈ Wh per 8-hour night |
|---|---|---|---|
| SlimLine tube, no humidifier | 0.93 A | ~11 W | ~89 Wh |
| SlimLine tube, humidifier at 4 | 2.30 A | ~28 W | ~221 Wh |
| SlimLine tube, humidifier at 8 | 3.66 A | ~44 W | ~351 Wh |
| ClimateLine Air heated tube at 30°C + humidifier at 8 | 4.92 A | ~59 W | ~472 Wh |
Watts = amps × 12 V; watt-hours = watts × 8 hours. Our arithmetic from ResMed’s published current. Higher pressure settings draw more: with no humidifier at 20 cmH₂O, the same table lists 1.66 A (~20 W).
That’s a 5x spread between bare and fully heated. In a summer outage, many people choose to skip humidification for a night or two, but whether that’s okay for you is a question for your doctor or supplier, not us.
Two cautions from the same ResMed guide: it recommends DC-to-DC converters over inverters, and when an inverter is used it recommends a pure sine wave unit rated at least 300 W continuous / 500 W surge. It also warns that some older ResMed heated humidifiers must not be used with inverters at all. If you’ll use a power station’s AC outlets, look for “pure sine wave” on its spec page or in its manual.
AirSense 11 owners: ResMed’s battery guide (dated 2018) doesn’t cover the AirSense 11, and ResMed’s own AirSense 11 FAQ says it doesn’t endorse any third-party battery solution for that machine and that it isn’t compatible with ResMed’s older Power Station II battery (ResMed AirSense 11 FAQ, Australian site, checked Oct 12, 2026). Ask ResMed or your supplier which DC options, if any, they support for your model in the U.S. Until then, plan on the AC figures.
Step 3: AC outlet or DC port?
A power station can feed a CPAP two ways:
- AC outlet + the machine’s normal adapter. This works with any CPAP and is what most people will do. It’s also the least efficient path: the station converts battery DC to AC, and the CPAP’s adapter converts it back to DC. We assume 85% inverter efficiency across the site (methodology), and an idling inverter can draw extra power that we don’t model.
- 12 V DC (car-style outlet) + the manufacturer’s DC cord. It skips the inverter, so it stretches runtime. Philips says to use only its own DC power cord and battery adapter cable, because other setups may damage the device (DreamStation 2 manual). ResMed sells DC converters for some models. Check that your station’s 12 V port amp rating covers the cord’s draw.
Step 4: Estimated nights per charge
Three planning scenarios. Each assumes 8 hours a night, a new battery, and that the station’s full rated capacity is usable. Real results vary with pressure, humidity settings, room temperature, and mask leak.
| Model | Rated capacity | AirSense 11 via AC (56.1 W typical) 528 Wh from the battery per night | AirSense 10 via AC (53 W typical) 499 Wh from the battery per night | AirSense 10 via DC, no humidifier (~11 W) 89 Wh from the battery per night (upper bound) |
|---|---|---|---|---|
| EcoFlow DELTA 3 Plus | 1,024 Wh | ~1.9 nights | ~2.1 nights | ~11 nights |
| Anker SOLIX C1000 Gen 2 | 1,024 Wh | ~1.9 nights | ~2.1 nights | ~11 nights |
| Jackery Explorer 1000 v2 | 1,070 Wh | ~2.0 nights | ~2.1 nights | ~12 nights |
| BLUETTI AC180 | 1,152 Wh | ~2.2 nights | ~2.3 nights | ~13 nights |
| Jackery Explorer 2000 v2 | 2,042 Wh | ~3.9 nights | ~4.1 nights | ~23 nights |
| Jackery HomePower 3600 Plus | 3,584 Wh | ~6.8 nights | ~7.2 nights | ~40 nights |
| Anker SOLIX F3800 Plus | 3,840 Wh | ~7.3 nights | ~7.7 nights | ~43 nights |
| EcoFlow DELTA Pro 3 | 4,096 Wh | ~7.8 nights | ~8.2 nights | ~46 nights |
Power station specs come from each manufacturer’s official page, last checked October 10, 2026. See the cost per kWh comparison for prices and links to the spec pages.
Which size makes sense?
- One or two nights, CPAP only: a ~1 kWh unit, such as the BLUETTI AC180, Jackery Explorer 1000 v2, EcoFlow DELTA 3 Plus, or Anker SOLIX C1000 Gen 2, covers about two full-featured nights on AC, and many more on DC with the humidifier off.
- CPAP plus the fridge: a fridge rated at 600 kWh/yr needs roughly 1,900 Wh per day from the battery (fridge guide), so the CPAP is the small part. Look at ~2 kWh and up, such as the Jackery Explorer 2000 v2, or plan to recharge daily.
- Multi-day outages: pair any size with solar. About 400 W of panels can refill a ~1 kWh station in roughly 3 peak sun hours (solar recharge guide). If you’re also considering a gas generator, read the CO rules in power station vs gas generator first. Generators never belong indoors.
If you use oxygen concentrators or other power-dependent equipment as well, size those separately with your provider. A concentrator draws far more than a CPAP.
Before the next storm
- Do a test night before you need it. Run the CPAP from the fully charged power station on a normal night and see how much charge is left in the morning. That beats any table, including ours.
- Keep the station charged. Top it up before forecast storms, and follow the station’s manual for storage charging. Our free power outage prep checklist covers the rest of a household outage plan.
- Tell your utility and fire department. FDA’s tips say to let them know you have a medical device that needs power. Many utilities run medical-priority programs, so ask yours.
- Check settings when power returns. FDA notes that medical devices often reset to defaults after a power interruption.
- Keep it dry. Don’t plug in a wet cord or device (FDA).
- Never backfeed a house circuit to run a bedroom outlet. Plug the CPAP straight into the station. For hardwired circuits, see the transfer switch and generator inlet guide.
FDA also publishes a fill-in power outage booklet for home medical devices (checked Oct 12, 2026) with questions to ask your manufacturer, such as “Can my device use batteries?” and “What happens if I lose power in the middle of a treatment?”
Sources
- ResMed, AirSense 11 user guide (Americas), technical specifications (checked Oct 12, 2026)
- ResMed, AirSense 10 user guide (Americas), technical specifications (checked Oct 12, 2026)
- ResMed, Battery Guide, doc. 198103/7 (2018-08), Air10 series converter table and inverter guidance (checked Oct 12, 2026)
- ResMed, AirSense 11 FAQs (Australia) (checked Oct 12, 2026)
- Philips, DreamStation 2 Advanced user manual (checked Oct 12, 2026)
- U.S. FDA, FDA Offers Tips about Medical Devices and Natural Disasters and home-use device power outage booklet (checked Oct 12, 2026)
- Power station specs: official pages for each model in our dataset, listed in the cost per kWh comparison (checked Oct 10, 2026)